It was reported that during an inspection, a system notice was received indicating that the safety system detected blood in the catheter handle, the injection was stopped and the vacuum disabled.Additionally, a system notice was received indicating that the refrigerant delivery path was obstructed, a system notice was received indicating that there was a problem with the refrigerant port and a system notice was received indicating that the safety system detected fluid in the catheter and stopped the injection.It was observed that there was blood in the console's coaxial umbilical connector.It was further reported that it is unknown when or how the blood got into the coaxial port.The outcome of this case is unknown. the data files were reviewed and confirmed the system notices.Test ablations were performed and no issues occurred.It was recommended to use the back-up console until service can be performed.Service was performed.No patient complications have been reported as a result of this event.
|
Product event summary: the afapro28 balloon catheter with lot number 15536 was returned and analyzed.External visual inspection of the balloon segment showed blood/fluid inside the balloon.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for five applications on the reported event date.During functional testing, the console terminated the application and triggered system notice 50005 "the safety system detected fluid in the catheter and stopped the injection." pressure testing and inspection was performed on the sub-components of the balloon, handle, and shaft segments.During inspection and pressure testing of the shaft segment, a guide wire lumen kink and breach was observed 0.74 inches proximal to the catheter tip.During inspection of the handle segment, blood / liquid was observed inside handle.In conclusion, the reported visible blood was confirmed through testing.The balloon catheter failed the returned product inspection due to the guide wire lumen kink and breach.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|