Catalog Number AA6120
Device Problem Unintended Deflation (4061)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/06/2023
Event Type  malfunction  
Event Description
According to the available information, on day 1 of the prostatectomy the catheter fell out and it was observed the balloon was deflated.
 
Manufacturer Narrative
After receiving this complaint, we searched for other complaints, and we didn¿t find any other complaint on the lot number 9261070.Our subcontractor made a documentary investigation which revealed no anomaly recorded during production, but this issue could be caused by a raw material issue.Checking the quality databases revealed a connection with the described defect with capa -000152 "balloon issues on folysil and silicone prostatic catheters".The trend for deflation issue on folysil catheters is specifically followed.Similar case study was done based on same item number and same defect [deflates] from november 2019 to november 2023: 6 similar cases were found.A risk management framework evaluation was performed and concluded that residual risks are adequately controlled and reduced as far as possible, and the residual risks associated with the use of the product are acceptable when weighed against the benefits to the patient/user.It is concluded that the risks identified are still acceptable and considered as safe.
 
Event Description
According to additional available information, the balloon was tested (inflation/deflation) in order to insert the catheter and the integrity of the balloon was checked after deflation.There were no missing pieces.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLYSIL SILICONE CATHETER
Type of Device
INDWELLING URETHRAL DRAINAGE BALLOON CATHETER, NON-ANTIMICROBIAL
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer (Section G)
CMF-SARLAT
9 avenue edmond rostand
sarlat-la-caneda
FR  
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key18221826
MDR Text Key329167588
Report Number9610711-2023-00254
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K013174
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAA6120
Device Lot Number9226766_AA61201002
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/28/2023
Supplement Dates Manufacturer Received10/20/2015
Supplement Dates FDA Received05/29/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured06/22/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-