Model Number V60
Device Problems Defective Alarm (1014); Insufficient Information (3190)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/17/2023
Event Type  malfunction  
Manufacturer Narrative
E: (b)(6).
 
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had a startup error.There was no patient involvement when the issue occurred.No patient or user harm reported.Investigation is ongoing.
 
Manufacturer Narrative
H10: a philips service engineer (pse) evaluated the device, and reported that the device had an error, and that the alarm light was faulty.The pse restarted the device, and the problem persisted after reconnecting the alarm light switch.The se noted that the power switch overlay was replaced to resolve the issue.The device was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18222702
MDR Text Key329575398
Report Number2518422-2023-32176
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public00884838025776
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberV60
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/17/2023
Initial Date FDA Received11/28/2023
Supplement Dates Manufacturer Received11/24/2023
Supplement Dates FDA Received11/30/2023
Date Device Manufactured03/06/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-