MAUDE Adverse Event Report: MAILLEFER INSTRUMENTS HOLDING SARL START-X TIP EMS INSERT 1; SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAILLEFER INSTRUMENTS HOLDING SARL START-X TIP EMS INSERT 1; SCALER, ULTRASONIC Back to Search Results
Catalog Number A066100000100
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
In this event it is reported that start-x tip ems insert 1 tip broke during use.The broken parts have been retrieved from the patient's mouth; no injury resulted.
 
Manufacturer Narrative
While no serious injury resulted in this event, there has been a previous report received with a similar device where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
 
Manufacturer Narrative
Summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused insert is available for evaluation.Nothing unusual to report was found during dhr review (batch #1814213).Information given regarding technique is only partial, we cannot rule on its compliance with maillefer's recommendations.Root causes are not identified.We will track this kind of event and monitor the trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
START-X TIP EMS INSERT 1
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ  CH-1338
Manufacturer (Section G)
MAILLEFER INSTRUMENTS HOLDING SARL
ch. du verger 3
ballaigues vaud CH-13 38
SZ   CH-1338
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key18228805
MDR Text Key329260741
Report Number8031010-2023-00878
Device Sequence Number1
Product Code ELC
Combination Product (y/n)N
Reporter Country CodeIS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberA066100000100
Device Lot Number1814213
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date11/24/2023
Initial Date Manufacturer Received 11/24/2023
Initial Date FDA Received11/29/2023
Supplement Dates Manufacturer Received11/24/2023
Supplement Dates FDA Received11/30/2023
Was Device Evaluated by Manufacturer? No
Type of Device Usage A
Patient Sequence Number1
-
-