• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG ELECTRODE, BIPOLAR; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG ELECTRODE, BIPOLAR; ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE-USE Back to Search Results
Model Number 011250-01
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
The affected device has been requested for investigation by the manufacturer.Device was not yet returned for investigation.The event is filed under internal karl storz complaint id: (b)(4).
 
Event Description
It was reported that the loops broke off while inside the patient and had to be retrieved.Due to the required surgical.Intervention to retrieve the pieces broken of the loops the event is reportable.
 
Manufacturer Narrative
As the device in question was not returned by the customer, an investigation on the product itself could not be performed.However, the available information has been reviewed.The investigation was completed on 2024-01-05.Based on the available information, the failure description and our experience, we suspect a mechanical overload of the cutting loop which led to the reported breakage.Due to the fact, that the product has not been sent back, a final determination of the root cause is not possible.Furthermore, the device history records have been checked and found to be according to the specification, valid at the time of production.There is no indication for a material-, manufacturing- or design-related failure.The root cause of the reported issue can most likely be traced back to a usage-related failure.
 
Manufacturer Narrative
The customer has returned the product back for analysis on may 15, 2024.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECTRODE, BIPOLAR
Type of Device
ENDOSCOPIC ELECTROSURGICAL ELECTRODE, BIPOLAR, SINGLE-USE
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM  78532
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
tuttlingen, 78532
GM   78532
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
MDR Report Key18231057
MDR Text Key330141925
Report Number9610617-2023-00385
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number011250-01
Device Catalogue Number011250-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received11/29/2023
Supplement Dates Manufacturer Received01/05/2024
Supplement Dates FDA Received01/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-