SHIRAKAWA OLYMPUS CO., LTD. VISERA CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-V2
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2023
Event Type  malfunction  
Event Description
The customer reported to olympus that the visera cysto-nephro videoscope was blown due to the pressure cap not being fully secured.The issue was found during reprocessing.A delay was reported but the patient was not under sedation and the procedure was completed with another device.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation was confirmed.The device evaluation found the bending section cover was ruptured.In addition to the reportable malfunction, the following were noted: the adhesive at the distal end was chipped and lifting.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the pressure cap was not completely secured, the problem was caused by user's mishandling.The contents of the device's instruction manual regarding reprocessing were reviewed and confirmed to contain a description to prevent this event.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISERA CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18231901
MDR Text Key329287180
Report Number3002808148-2023-13470
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339431
UDI-Public04953170339431
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K221683
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/03/2023
Initial Date FDA Received11/29/2023
Supplement Dates Manufacturer Received12/15/2023
Supplement Dates FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/22/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-