• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. DUODENOFIBERSCOPE FOR ULTRASONIC SURVEY; ULTRASOUND GASTROVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. DUODENOFIBERSCOPE FOR ULTRASONIC SURVEY; ULTRASOUND GASTROVIDEOSCOPE Back to Search Results
Model Number TGF-UC260J
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned and evaluated, and the customer¿s allegation was confirmed.In addition to the malfunction documented in b5, the additional evaluation findings are as follows: due to deformation of u/d (up/down) knob, water tightness was lost, due to a pinhole on ch (channel) tube, water tightness was lost, due to wear of angle wire, bending angle in up direction did not meet the standard value, due to wear of angle wire, the play of u/d knob was out of the standard value, adhesive on a-rubber was detached, and the ultrasonic transducer had corrosion.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, that the duodenofiberscope had an air leak.The device was returned for evaluation.During the device evaluation, it was found that the probe was chipped.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation (h6/10).A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause could not be conclusively determined.The following information from the instructions for use (ifu) pertains to this event: "warning do not strike, hit, or drop the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector.Also, do not bend, pull, or twist the endoscope¿s distal end, insertion tube, bending section, control section, universal cord, or endoscope connector with excessive force.The endoscope may be damaged and could cause patient injury, burns, bleeding, and/or perforations.It could also cause parts of the endoscope to fall off inside the patient." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUODENOFIBERSCOPE FOR ULTRASONIC SURVEY
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18233951
MDR Text Key329915355
Report Number3002808148-2023-13484
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTGF-UC260J
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/07/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received11/29/2023
Supplement Dates Manufacturer Received12/06/2023
Supplement Dates FDA Received12/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/17/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-