Model Number 5826
Device Problems Pacemaker Found in Back-Up Mode (1440); Therapy Delivered to Incorrect Body Area (1508)
Patient Problem Discomfort (2330)
Event Date 11/16/2023
Event Type  malfunction  
Event Description
It was reported that the patient presented for monopolar electrosurgery when the pacemaker suddenly went into backup vvi operation and the patient experienced left pectoral muscle stimulation.The device was reprogrammed to normal mode.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR XL DR
Type of Device
Pulse generator, permanent, implantable
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18233991
MDR Text Key329465348
Report Number2017865-2023-93502
Device Sequence Number1
Product Code NVZ
UDI-Device Identifier05414734501620
UDI-Public05414734501620
Combination Product (y/n)N
Reporter Country CodeRS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Physician
Type of Report Initial
Report Date 11/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Model Number5826
Device Lot Number4336210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received11/29/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/08/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-