Summary of investigation: there are two previous complaints of this code-batch, one of them regarding the same issue (closed as not confirmed after analysis).We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample, only a picture of an open sample.However, without closed samples and/or defective samples a proper analysis cannot be performed.Batch manufacturing record: reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Conclusion root cause analysis: it has not been possible to determine the root cause as no samples are received.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done, and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Corrective measures: actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
|