• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN SURGICAL, S.A. DAFILON BLUE 4/0 (1.5) 75CM DS19; OTHER SUTURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN SURGICAL, S.A. DAFILON BLUE 4/0 (1.5) 75CM DS19; OTHER SUTURE Back to Search Results
Model Number C0935204
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/13/2023
Event Type  malfunction  
Manufacturer Narrative
If additional information becomes available a follow-up report will be submitted.
 
Event Description
It was reported an issue with dafilon suture.The customer reported that during surgery, the thread is detached close to the needle.No further information received.
 
Manufacturer Narrative
Summary of investigation: there are two previous complaints of this code-batch, one of them regarding the same issue (closed as not confirmed after analysis).We manufactured and distributed in the market (b)(4) units of this code-batch.There are no units in our stock.We have not received any sample, only a picture of an open sample.However, without closed samples and/or defective samples a proper analysis cannot be performed.Batch manufacturing record: reviewed the batch manufacturing record, this product had a normal process and the results during the process fulfill usp/ep and b.Braun surgical requirements.Conclusion root cause analysis: it has not been possible to determine the root cause as no samples are received.Final conclusion: without samples we are not in position of studying if the affected product does not fulfil the specifications.In consequence, a proper analysis cannot be done, and the case is not confirmed due to lack of evidence.Nevertheless, we take note of this incidence and if any sample is received in the future, we will re-open the case and analyse it.Please note that when no samples are received our analysis is very limited.Corrective measures: actions on distributed product of this reference/batch: based on the conclusion derived from investigation, it is not required to make actions in distributed product.Corrective/preventive actions: according to our internal procedures, there is no need to establish corrective or preventive actions.Nevertheless, this complaint is recorded for trending analysis to assess if actions are needed in the future.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAFILON BLUE 4/0 (1.5) 75CM DS19
Type of Device
OTHER SUTURE
Manufacturer (Section D)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP  08191
Manufacturer (Section G)
B. BRAUN SURGICAL, S.A.
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
Manufacturer Contact
martina laporte
carretera de terrassa 121
rubí, barcelona 08191
SP   08191
MDR Report Key18235890
MDR Text Key329319860
Report Number3003639970-2023-00419
Device Sequence Number1
Product Code GAR
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K151165
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC0935204
Device Catalogue NumberC0935204
Device Lot Number622366
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/07/2023
Initial Date FDA Received11/30/2023
Supplement Dates Manufacturer Received01/08/2024
Supplement Dates FDA Received01/12/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
-
-