BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problem
Material Integrity Problem (2978)
|
Patient Problem
Device Overstimulation of Tissue (1991)
|
Event Date 11/10/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that during the implant of this cardiac resynchronization therapy pacemaker (crt-p), once placed into the pocket, bipolar right ventricular (rv) pacing resulted in chest muscle stimulation that was visually observed through from this device header.This crt-p had been taken out of storage mode today.Unipolar testing was fine with no stimulation.Rv lead and impedance testing showed normal measurements.The right atrial (ra) lead was placed into the rv port for testing purposes in which the same issue was observed.Rv only pacing was attempted and was not biventricular (bv) pacing at the time.True bipolar on lv lead was observed.Bipolar pacing was attempted on the pacing system analyzer (psa) with no issues.The psa ring unipolar did not reproduce it.Muscle stimulation was also recreated with the rv lead in the ra port, noting that both ports of this crt-p header resulted in the stimulation.Technical services (ts) noted this was likely not a lead issue as the leads were tested fine with the psa.Ultimately, the physician decided to implant a different device.This new device remains in service.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
This device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination identified no anomalies.Functional testing was undertaken and the device did not perform as expected.Additionally, the device was found to be operating in safety mode.The device case was removed to facilitate inspection and testing of the internal components.High powered visual inspection found that the battery insulation liner did not completely insulate the battery, resulting in the battery case making contact with the device case.This resulted in the reported clinical observations and the reversion to safety mode operation.Correction to b1 to change to product problem and removed b2 selections.
|
|
Event Description
|
It was reported that during the implant of this cardiac resynchronization therapy pacemaker (crt-p), once placed into the pocket, bipolar right ventricular (rv) pacing resulted in chest muscle stimulation that was visually observed through from this device header.This crt-p had been taken out of storage mode today.Unipolar testing was fine with no stimulation.Rv lead and impedance testing showed normal measurements.The right atrial (ra) lead was placed into the rv port for testing purposes in which the same issue was observed.Rv only pacing was attempted and was not biventricular (bv) pacing at the time.True bipolar on lv lead was observed.Bipolar pacing was attempted on the pacing system analyzer (psa) with no issues.The psa ring unipolar did not reproduce it.Muscle stimulation was also recreated with the rv lead in the ra port, noting that both ports of this crt-p header resulted in the stimulation.Technical services (ts) noted this was likely not a lead issue as the leads were tested fine with the psa.Ultimately, the physician decided to implant a different device.This new device remains in service.No adverse patient effects were reported.
|
|
Search Alerts/Recalls
|
|
|