MEDTRONIC NAVIGATION, INC (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, M
|
Back to Search Results |
|
Model Number BI70000027120 |
Device Problems
Application Program Problem (2880); Communication or Transmission Problem (2896); Radiation Output Problem (4026)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/09/2023 |
Event Type
malfunction
|
Event Description
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that when attempting to perform and automatic registration spin the system produced a "scan could not be transferred at this time" error message.The representative attempted to manually push the scans from the mobile view station (mvs) to the navigation system and would not allow transfer.The representative attempted to re-spin and the system functioned as intended.There was no patient involvement.
|
|
Manufacturer Narrative
|
H3) onsite functional and visual examination was performed by a manufacturer representative.The system passed a system checkout and was determined to be operational.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received.There was a connection between the imaging system and navigation system.The communication box was green before, during, and after the spin.The images didn¿t transfer automatically and when the field representative (rep) attempted to push the images to the navigation system, they received an error.Another spin had to be completed to move forward with surgery.There was a patient present.
|
|
Manufacturer Narrative
|
H2) please see section b5 for additional information.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|