RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/21/2023 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint on the v60 ventilator indicating that the device displayed an occlusion error, and the blower was not running.The device was reported to be outside of use at the time of the reported problem.There was no patient harm reported.The remote service engineer (rse) was informed by the customer that the customer biomedical engineer (bme) disassembled the blower and advised that there was internal contamination.The rse advised the bme to order the blower assembly, rubber boot kit, and tubing kit.The customer confirmed that installing a new blower assembly resolved the device issue.
|
|