Model Number URF-V3 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/19/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus that during routine microbiological testing, the uretero-reno videoscope tested positive for 4 colony forming units (cfus)/100ml of micrococcus, neisseria, and staphylococcus epidermidis.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
The olympus scope was sent to an independent laboratory for culture testing and results are still pending.The investigation is ongoing and follow up with the customer is currently being performed.After culture testing, the device will be evaluated.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on results of third-party testing, the device evaluation and the legal manufacturer's final investigation.Despite good faith attempts the user cleaning disinfection and sterilization (cds) processes were not shared.Olympus provided the following result of the culture test, performed at the third-party labs: sampling date: (b)(6) 2023.Sampling from: all channels cfu: <1cfu.Bacterial identification: n/a.The device was evaluated where no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The following is included in the device ifu: "an insufficiently reprocessed endoscope and/or accessory may pose an infection control risk to the patients and/or operators who touch them." olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|