Model Number 865350 |
Device Problem
No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/20/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information and the complaint is still under investigation.A follow-up report will be submitted upon completion of the investigation.
|
|
Event Description
|
It was reported the device had no audio.It is unknown if the device was in clinical use at the time of the event, no adverse event or patient harm was reported.
|
|
Manufacturer Narrative
|
It was reported the device had no audio.The device was not in clinical use at the time of the event.The device was sent to a philips authorized repair facility that performed diagnostic/functional testing. results of the functional testing indicate that there was no speaker sound at the start up test and the speaker was defective.The repair facility replaced the speaker.Based on the information available and the testing conducted, the cause of the reported problem was a defective speaker. the investigation concludes that no further action is required at this time.
|
|