Model Number IPN041365 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It was reported that " during induction of a child, no capnia value/measurement was displayed on the ventilator because the filter was blocked.Patient was reported as fine post the procedure.No reported patient harm.Associated to 8040412-2023-00397.
|
|
Event Description
|
It was reported that "during induction of a child, no capnia value/measurement was displayed on the ventilator because the filter was blocked.Patient was reported as fine post the procedure.No reported patient harm.Associated to 8040412-2023-00397.
|
|
Manufacturer Narrative
|
(b)(4).The device was not returned for investigation; however, the manufacturing site reports "this defect had been acknowledged as supplier related issue.The mentioned product housing was supplied by our supplier, and hence supplier corrective action request (scar) and nc has been issued, root cause analysis and identified corrective actions will be implemented through this scar.Assembly process will be conducted in our manufacturing site for the product after receive the part item from supplier.".
|
|