• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40, INTERNATIONAL; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40, INTERNATIONAL; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number 1111169
Device Problems Electrical /Electronic Property Problem (1198); Device Emits Odor (1425)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/17/2019
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator alarming error code and giving off an electrical odor.There was no harm or injury reported.During the evaluation of the device at the manufacturer's service center the error code was confirmed.No service action was required for the error code.The electrical odor fault was not confirmed but the odor was caused by tar and nicotine.The blower was contaminated with tar and the flow path was yellowing which was caused by the nicotine.The unit was scrapped.
 
Manufacturer Narrative
H3 other text : serviced by third party service center.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40, INTERNATIONAL
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18246252
MDR Text Key329482815
Report Number2518422-2023-32769
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K121623
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1111169
Device Catalogue Number1111169
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/17/2019
Initial Date FDA Received12/01/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/10/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-