Model Number ENSITE-AMP-02 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/23/2023 |
Event Type
malfunction
|
Event Description
|
During the atrial tachycardia procedure, output issues which troubleshooting measures did not resolve, resulted in the cancellation of the procedure.While in the left atrium, it was noted that the error message, problem with prs-p sensores' displayed on the ensite x.There was no noticeable patient movement but the system indicated that the sensors had lost their initial position.It was attempted to reposition the sensors but this did not resolve the issue.The physician decided to cancel the procedure due to the unclear impact to catheter location and potential for patient harm.Cardioversion was performed to terminate the atrial tachycardia with no adverse consequences to the patient.
|
|
Manufacturer Narrative
|
The results of the investigation were inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident could not be determined.
|
|