ST. JUDE MEDICAL, INC. ENSITE¿ X AMPLIFIER; COMPUTER, DIAGNOSTIC, PROGRAMMABLE Back to Search Results
Model Number ENSITE-AMP-02
Device Problem Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2023
Event Type  malfunction  
Event Description
During the atrial tachycardia procedure, output issues which troubleshooting measures did not resolve, resulted in the cancellation of the procedure.While in the left atrium, it was noted that the error message, problem with prs-p sensores' displayed on the ensite x.There was no noticeable patient movement but the system indicated that the sensors had lost their initial position.It was attempted to reposition the sensors but this did not resolve the issue.The physician decided to cancel the procedure due to the unclear impact to catheter location and potential for patient harm.Cardioversion was performed to terminate the atrial tachycardia with no adverse consequences to the patient.
 
Manufacturer Narrative
The results of the investigation were inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information received, the cause of the reported incident could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENSITE¿ X AMPLIFIER
Type of Device
COMPUTER, DIAGNOSTIC, PROGRAMMABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18248547
MDR Text Key329504162
Report Number2184149-2023-00230
Device Sequence Number1
Product Code DQK
UDI-Device Identifier05415067034755
UDI-Public05415067034755
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K231415
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberENSITE-AMP-02
Device Lot Number9129128
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/23/2023
Initial Date FDA Received12/01/2023
Supplement Dates Manufacturer Received12/05/2023
Supplement Dates FDA Received12/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/14/2023
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-