The customer reported to olympus that during routine microbiological testing, the cysto-nephro videoscope tested positive for > 100 colony forming units (cfus) of microorganisms.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
Updated: d8, d9, h3, h4, h6, h10 this report is being supplemented to provide additional information based on results of third-party testing, the customer provided cleaning disinfection and sanitization (cds) practices, the device evaluation, and the legal manufacturer's final investigation.Additional information received from the customer regarding the facility cleaning disinfection and sterilization practices (cds) stated that there were no deviations, deficiencies or concerns during device reprocessing.The device was evaluated and no abnormalities were found that could have led to the positive culture.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing, however, when olympus culture tested after reprocessing in accordance with instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The event can be prevented by following the instructions for use below: - cysto-nephro videoscope olympus cyf-vh olympus cyf-vhr reprocessing manual olympus will continue to monitor field performance for this device.
|