Model Number CYF-VH |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/29/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The hygiene microbiological investigation report indicated the channels of the scope were cultured and found less than 1 cfu of any microbes.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.The customer provided the cleaning sterilization and disinfection (cds) processes performed at the user facility.The customer did confirm that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The customer did not provide the specific steps taken during the cleaning sterilization and disinfection (cds) process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Event Description
|
The customer reported to olympus, during reprocessing, the cysto-nephro videoscope initially tested positive for 1 colony forming unit (cfu) of niallia circulans on (b)(6), 2023.The device was retested on (b)(6), 2023, and the test result was positive for 2 cfus of metabacillus spp.And an unidentifiable bacterium.The device was tested again on (b)(6) 2023.The third test result was positive for 6 cfus of micrococcus luteus, moraxella osloensis, bacillus sp., paenibacillus provencensis, and agrobacterium radiobacter.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.The device was evaluated where no abnormalities were found that could have led to the positive culture.The following defects were noted: bending section rubber adhesive separation, biopsy channel specifications.Forceps movement (failure), paint peeling on connecting tube, channel mount wear, mouthpiece defects, case cracked, and cracked plug unit.However, these defects alone are not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The cyf-vh reprocessing method is described in the "cysto-nephro videoscope olympus cyf-vh olympus cyf-vhr reprocessing manual".This may prevent the indicated phenomenon.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|