Catalog Number 110029136 |
Device Problem
Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
It was reported that during a procedure, the surgeon noted the mini baseplate reamer was dull and did not want to use it.A second device was used to complete the surgery.There were no patient consequences as a result of the event.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.H6: proposed component code - mechanical (g04)- drill.Visual examination of the returned product identified the cutting edges of the reamer are worn and dull.The device shows heavy signs of repeated use.Medical records were not provided.Review of the device history record(s) identified no deviations or anomalies during manufacturing.A definitive root cause cannot be determined.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Search Alerts/Recalls
|