Situation - aerobic blood culture bottles are coming to 9wp damaged.Sterile process is not maintained with these bottles.Background - staff on 9wp noticed multiple blue top blood culture bottles are being found damaged with the sterile top being indented.Staff will remove the cap off these devices and noticed the gray center is inverted.Analysis- these bottles should not be used for blood cultures as we do not know if there was a breach in sterile process, needles are not able to be used to inject blood into the bottles due to not having a flat service.Lot number 3102697.Exp date [redacted date].Not sure if other lot numbers are involved.Reports this happening over the weekend as well but staff did not save bottles or record lot numbers.Recommendation - value analysis and/or materials management collect device and determine what needs to happen further.Manufacturer response for blood culture bottles, aerobic blood culture bottle (per site reporter).[redacted date] - emailed bd complaints.Bd case (b)(4).
|