ZIMMER BIOMET, INC. BONE SCREW SELF-TAPPING 6.5 MM DIA. 25 MM LENGTH; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problems Difficult to Insert (1316); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/24/2023
Event Type  malfunction  
Event Description
It was reported that during a procedure the screw penetrated the screw hole.The head section of the screw was deformed.Another screw was used to complete the procedure.There is no additional information available at the time of this report.
 
Manufacturer Narrative
(b)(4).D10: 00-6202-060-20, item name: tm acet shell 60mm multi, lot# 63307782.G2: foreign: japan.Multiple mdr reports were filed for this event, please see associated reports 0001822565 - 2023 - 03389.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
There is no update to the prior event description provided.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4 g3 g6 h2 h6 h10 visual examination of the returned product identified hex shows nicks and gouges from use.Outer head diameter has a lip formed from screw penetration through the shell causing deformation.Distal threads and taper below head show witness marks from insertion.The cup was not returned.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BONE SCREW SELF-TAPPING 6.5 MM DIA. 25 MM LENGTH
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18255966
MDR Text Key329599852
Report Number0001822565-2023-03388
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K934765
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00625006525
Device Lot NumberJ7358568
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/07/2023
Initial Date FDA Received12/04/2023
Supplement Dates Manufacturer Received02/14/2024
Supplement Dates FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/07/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
PLEASE SEE H10.
Patient SexPrefer Not To Disclose
-
-