ANGIODYNAMICS 400 MICRON FIBER PROCEDURE KIT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number EVLTPVAK
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
It was reported that the device hs yet to be returned to the manufacturer for evaluation.The results of the investigation will be sent via a follow up medwatch.Reference (b)(4).
 
Event Description
An end user reported an issue with a pvak -- 400 micron perforator and accessory vein ablation kit.While unpacking the device, the tip was found to have detached from the fiber.The following procedure was completed with another same device and the patient did not experience any adverse effects, harm, or require medical intervention as a result of this incident.
 
Manufacturer Narrative
The customer's reported complaint description of "fiber was fractured and detached" was not confirmed due to no sample being returned.Without receiving a sample for evaluation the root cause could not be determined.Manufacturing personnel 100% visually inspect devices during the packaging process.This type of fiber kink/detached damage would be noticed prior to shipment.A review of the device history records was performed for the indicated lots for any deviations related to the reported failure mode of the complaint.The review confirms that the lots met all material, assembly and performance specifications; i.E.No ncr associated with reported failure mode.Labeling review: the instructions for use (14601411-01), which is supplied to the end user with this catalog number, states: "prior to and during use, avoid damaging the fiber by striking, stressing, or excessive bending.Do not coil the fiber tighter than a diameter of 20cm.Clinical safety and effectiveness data is not available for other fiber tip designs and diameters.Prior to and during use, avoid bending the introducer sheath and dilator as this can cause kinks and damage".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
400 MICRON FIBER PROCEDURE KIT
Type of Device
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key18256646
MDR Text Key329608327
Report Number1319211-2023-00087
Device Sequence Number1
Product Code GEX
UDI-Device IdentifierH787EVLTPVAK5
UDI-PublicH787EVLTPVAK5
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K041957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberEVLTPVAK
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/13/2023
Initial Date FDA Received12/04/2023
Supplement Dates Manufacturer Received12/27/2023
Supplement Dates FDA Received01/09/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial