• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC CAVITON 300 SERIES G310SCLR(DNA); SCALER, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC CAVITON 300 SERIES G310SCLR(DNA); SCALER, ULTRASONIC Back to Search Results
Catalog Number G310SCLR
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Since an overheating insert could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803 the device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
While the customer was using a cavitron 300 series g310 they allege that the unit gets very hot to the touch and the cavitron tip has burnt.No injury was reported from the alleged event.
 
Manufacturer Narrative
The st is worn leaking water from the sides, sides of the display screen is cracked, heavy built up debris in water hose, solenoid and hpc wire harness, damaged power drive board, it delays activation when fc is pressed.Damaged and faulty parts have been replaced, unit have been calibrated and tested to factory's specs.No other faults found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAVITON 300 SERIES G310SCLR(DNA)
Type of Device
SCALER, ULTRASONIC
Manufacturer (Section D)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer (Section G)
DENTSPLY LLC
1301 smile way
york PA 17404
Manufacturer Contact
hannah seevaratnam
221 w. philadelphia st.
york, PA 17401
7178494593
MDR Report Key18256854
MDR Text Key329610680
Report Number2424472-2023-00056
Device Sequence Number1
Product Code ELC
UDI-Device IdentifierD00382703131
UDI-PublicD00382703131
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K150535
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberG310SCLR
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/04/2023
Supplement Dates Manufacturer Received11/30/2023
Supplement Dates FDA Received12/28/2023
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-