Model Number N/A |
Device Problem
Difficult to Insert (1316)
|
Patient Problem
Bone Fracture(s) (1870)
|
Event Date 11/09/2023 |
Event Type
Injury
|
Event Description
|
It was reported that during a bone marrow expansion, the stem was unable to be implanted all the way resulting in a bone fracture, increasing the patient's leg length and increasing the pain in the patient.Because of the femoral fracture, the use of the trochanteric claw and cable was necessary.There is no additional information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).G2: foreign: china.The device will not be returned for analysis as it remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: h6 component code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Complaint confirmed based on the evaluation of the provided radiograph.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: sizing is appropriate.Femoral component does not appear to be fully seated.Lesser trochanter fracture.No signs of loosening, wear, radiolucency.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
There is no update to the prior event description provided.
|
|
Search Alerts/Recalls
|