• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ARCOS 14X150MM SPL TPR DIST; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ARCOS 14X150MM SPL TPR DIST; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem Bone Fracture(s) (1870)
Event Date 11/09/2023
Event Type  Injury  
Event Description
It was reported that during a bone marrow expansion, the stem was unable to be implanted all the way resulting in a bone fracture, increasing the patient's leg length and increasing the pain in the patient.Because of the femoral fracture, the use of the trochanteric claw and cable was necessary.There is no additional information available at the time of this report.
 
Manufacturer Narrative
(b)(4).G2: foreign: china.The device will not be returned for analysis as it remains implanted; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: h6 component code: mechanical (g04) - stem.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Complaint confirmed based on the evaluation of the provided radiograph.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: sizing is appropriate.Femoral component does not appear to be fully seated.Lesser trochanter fracture.No signs of loosening, wear, radiolucency.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCOS 14X150MM SPL TPR DIST
Type of Device
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18257978
MDR Text Key329621856
Report Number0001825034-2023-02820
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K090757
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number11-300814
Device Lot Number65886009
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/10/2023
Initial Date FDA Received12/04/2023
Supplement Dates Manufacturer Received01/03/2024
Supplement Dates FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-