It was reported that during the sleeve gastrectomy procedure after the first firing using green load, when they removed the reload, the second reload and when stapler was closed they could not insert jaws of stapler thorough the same 12 millimeter trocar.Due to a gap between anvil and cartridge stapler jaws which the reason it was not able to fit through.Another like device was used to complete the procedure.There were no patient consequences.Buttress was used.
|
(b)(4).Date sent: 12/4/2023.D4: batch # 176c81.Investigation summary: the product was returned to ethicon endo-surgery for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one ech60l device was returned with the anvil bent upwards and without reload present.As additional testing, the device was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples meet the staple release criteria.It is possible that the device was clamped over an excess of tissue causing the anvil to bend.Please reference the instruction for use for more information. as part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 176c81, and no non-conformances were identified.
|