• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC NONE; ENDOSCOPE PLUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC NONE; ENDOSCOPE PLUS Back to Search Results
Model Number 470057-08
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Insufficient Information (4580)
Event Date 10/25/2023
Event Type  malfunction  
Manufacturer Narrative
Intuitive surgical, inc.(isi) did receive a da vinci product to perform failure analysis (fa).Fa was able to confirm the reported complaint and found the 30-degree endoscope was missing screw/disc at the camera instrument adapter (ciad).Fa also found the paddleboard had mechanical damage, the housing was discolored, and the transmittance test failed.
 
Event Description
It was reported that during central processing, the 30-degree endoscope had a damaged connector.There was no report of patient involvement and no reported injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NONE
Type of Device
ENDOSCOPE PLUS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key18260852
MDR Text Key330247977
Report Number2955842-2023-20887
Device Sequence Number1
Product Code GCJ
UDI-Device Identifier00886874116555
UDI-Public(01)00886874116555
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K191736
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other
Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date 11/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number470057-08
Device Catalogue Number470057
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/08/2023
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received12/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-