• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number R1111177
Device Problems Charging Problem (2892); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/12/2023
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator inoperative condition occurred.A salesperson visited the customer and reviewed the device error logs.A ventilator inoperative alarm was not observed, but a check ac power supply alarm was listed.There was no harm or injury reported.The device was sent to a service center for evaluation.The ventilator inoperative condition was not duplicated.During evaluation and review of the device error logs, it was identified that error e-95 causing the device to reboot.The device main board was replaced to address the issue.The check ac power supply error was confirmed.The ac power adapter and ac power connector were evaluated, and no abnormality was found.The power cable harness was replaced as a preventative measure.The unit was confirmed to be working properly after replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40 SYSTEM SILVER
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18263769
MDR Text Key329663978
Report Number2518422-2023-33418
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121623
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberR1111177
Device Catalogue NumberR1111177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2023
Initial Date Manufacturer Received 06/12/2023
Initial Date FDA Received12/04/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-