RESPIRONICS, INC. BIPAP A40 SYSTEM SILVER; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number R1111177 |
Device Problems
Charging Problem (2892); Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/12/2023 |
Event Type
malfunction
|
Event Description
|
The manufacturer received information alleging a ventilator inoperative condition occurred.A salesperson visited the customer and reviewed the device error logs.A ventilator inoperative alarm was not observed, but a check ac power supply alarm was listed.There was no harm or injury reported.The device was sent to a service center for evaluation.The ventilator inoperative condition was not duplicated.During evaluation and review of the device error logs, it was identified that error e-95 causing the device to reboot.The device main board was replaced to address the issue.The check ac power supply error was confirmed.The ac power adapter and ac power connector were evaluated, and no abnormality was found.The power cable harness was replaced as a preventative measure.The unit was confirmed to be working properly after replacement.
|
|