G.5.The heater-cooler 16-02-95 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in japan.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
Livanova deutschland received a report that a heater-coolersystem 3t was have cooling problem during a procedure.Cooling started from 38 at 11:30, and it took 70 minutes at 12:40 to reach 18.In addition, despite the setting of 15, the temperature drops to 18, and during cooling, the temperature drops to 18, then rises once to 21 or 22, and then drops again, making cooling unstable and taking a long time.There is no problem with heating.There is no report of any patient injury.
|