LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER 3T 200V 50/60HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-95
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
G.5.The heater-cooler 16-02-95 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in japan.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova deutschland received a report that a heater-coolersystem 3t was have cooling problem during a procedure.Cooling started from 38 at 11:30, and it took 70 minutes at 12:40 to reach 18.In addition, despite the setting of 15, the temperature drops to 18, and during cooling, the temperature drops to 18, then rises once to 21 or 22, and then drops again, making cooling unstable and taking a long time.There is no problem with heating.There is no report of any patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER 3T 200V 50/60HZ
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 41037
GM  41037
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18264605
MDR Text Key329707653
Report Number9611109-2023-00612
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901136
UDI-Public(01)04033817901136(11)191024
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-95
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/07/2023
Initial Date FDA Received12/05/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/24/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-