• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. SOLTIVE PREMIUM SUPERPULSED LASER SYSTEM; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. SOLTIVE PREMIUM SUPERPULSED LASER SYSTEM; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number TFL-PLS
Device Problems Temperature Problem (3022); Explosion (4006)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/16/2023
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to olympus for evaluation.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the soltive premium superpulsed laser system fiber exploded at the connector.The issue was found during initial activation.The laser ureteroscopy procedure completed.There were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.Correction to b5, d4, d10, e3, g2 for information inadvertently left out of previous report.Correction to h6 problem code.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device for this complaint was not returned and a physical inspection could not be performed.Based on the results of the investigation, this failure appears to be the result of the laser fiber breaking.No device malfunction was identified with the laser console and the console appears to have operated as expected.The console provided energy per input from the footswitch.Olympus will continue to monitor field performance for this device.
 
Event Description
The therapeutic laser ureteroscopy was completed using a similar replacement device.There was a delay of less than 5 minutes to replace the subject device.The related patient identifiers are as follows: (b)(6) for the soltive premium superpulsed laser system.(b)(6) for the 150 micron tfl ball tip single use fiber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLTIVE PREMIUM SUPERPULSED LASER SYSTEM
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18269078
MDR Text Key329747920
Report Number3011050570-2023-00212
Device Sequence Number1
Product Code GEX
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K183647
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTFL-PLS
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received12/05/2023
Supplement Dates Manufacturer Received02/28/2024
Supplement Dates FDA Received03/05/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/07/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
150 MICRON SINGLE USE FIBER (LOT KR260888)
-
-