• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SUPER SHEATH INTRODUCER SHEATH; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SUPER SHEATH INTRODUCER SHEATH; INTRODUCER, CATHETER Back to Search Results
Model Number 28186
Device Problems Difficult to Remove (1528); Difficult to Advance (2920); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/14/2023
Event Type  malfunction  
Event Description
It was reported that the sheath was difficult to remove.A 6fr introducer sheath was selected for use in a percutaneous transluminal angioplasty (pta) procedure to treat an occluded vessel.During insertion of the sheath, there was difficulty advancing the sheath.When the sheath was removed, the tip was deformed.It was difficult to remove, and the tip caught on the vessel wall.A new sheath was selected for use to complete the procedure.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER SHEATH INTRODUCER SHEATH
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
MEDIKIT COMPANY LIMITED
13-2 yushima 1-chome
bunkyo-ku tokyo 113-0 034
JA   113-0034
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18270360
MDR Text Key329755710
Report Number2124215-2023-67150
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number28186
Device Catalogue Number28186
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/15/2023
Initial Date FDA Received12/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-