Model Number 96102 |
Device Problem
No Display/Image (1183)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/05/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
A spacelabs field service engineer (fse) reached out to the customer.The fse reported that the device was removed from use and will be repaired.A follow up report will be submitted in accordance with 21 cfr 803.56 when additional information becomes available.
|
|
Event Description
|
The customer reported that while in use, an xhibit central station had blank displays.Both local displays and remote displays were affected.There was no patient or user harm associated with this event.
|
|
Manufacturer Narrative
|
The unit was shipped to spacelabs for depot repair.A equipment service technician tested the unit and identified a corrupted hard drive.The hard drive was replaced and the reported problem was resolved.Cause of failure was a corrupted hard drive.
|
|