Model Number URF-V3R |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation and the customer¿s allegation was confirmed due to a hole in the bending section cover.The evaluation found a failed leak test even with bending section cover being taped.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus that the uretero-reno videoscope experienced an air water leakage form the distal end causing a failed leak test.The issue was found during reprocessing.The device was returned for evaluation.During the device evaluation, the bending section cover supporter pins were found to be lifted and sharp.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to cable support came off by forcible bending in opposite direction while bending section was not moving.The event can be detected by following the instructions for use (ifu) section: -inspection of the endoscope the event can be prevented by following the instructions for use (ifu) which state: -do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|