• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA III XENON LIGHT SOURCE Back to Search Results
Model Number CLV-190
Device Problems Unintended Collision (1429); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found a broken front panel, scope detection sensor failure, and a non-olympus lamp with more than 200 hours of use.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the evis exera iii xenon light source had connector issues, the output socket or the light guide connector was broken.The issue was found during preparation for use.There were no reports of patient harm.
 
Event Description
The customer clarified the event did not occur before a procedure, and a non-olympus device collided with the processor and damaged the outer layer.The procedures were completed with the same set of equipment.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over eight (8) years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be determined.However, it is likely that the scope detection is not functioning was due to a broken output socket or light-guide connector.The following information is stated in the instructions for use (ifu): warning.Never install a lamp that has not been approved by olympus.The use of a nonapproved lamp can cause damage to the light source and ancillary equipment, malfunction or a fire.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18279691
MDR Text Key329929827
Report Number3002808148-2023-13854
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/21/2023
Initial Date FDA Received12/06/2023
Supplement Dates Manufacturer Received12/22/2023
Supplement Dates FDA Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/08/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-