MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P7R
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/10/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer¿s allegation was not confirmed.The technician was unable to observe nor replicate the alleged lubrication leak.In addition, the following non-reportable malfunctions were noted during evaluation: the image on the olympus endoscope system had image guide brakeage and the centering is off.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Event Description
The customer reported to olympus that the uretero-reno fiberscope had foreign material falling from the channel and there was lubrication leaking out.The issue was observed during receipt inspection and noted as an out-of-box failure for use on an unknown procedure.There was no patient harm or user injury reported due to the event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on results of the investigation, the reported foreign material/lubricant leaking from the scope issue could not be reproduced or confirmed during the device evaluation; no device leakage was found.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18281523
MDR Text Key329890320
Report Number9610595-2023-18980
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K181451
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P7R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/15/2023
Initial Date FDA Received12/07/2023
Supplement Dates Manufacturer Received12/15/2023
Supplement Dates FDA Received01/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/11/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-