Model Number IV
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A healthcare professional reported with a description of inserter pleated.Additional information has been requested, received and stated the event occurred prior to surgery with no patient contact.The operation has been completed using the backup.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
One company model handpiece was returned for evaluation for the report of the injector pleated.A visual inspection of the intraocular lens (iol) handpiece injector was performed and was found to be nonconforming.The plunger was bent.A functional thread to barrel engagement check was performed and found to be conforming.Finally, a dimensional inspection for plunger position height was performed and was found nonconforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The evaluation does confirm the injector plunger has a bent plunger head resulting in a upward plunger position, which could result in the injector pleated.The root cause for the nonconforming plunger height is unknown.How and when how the plunger position became damaged cannot be determined from this evaluation.The most likely cause of a bent plunger head of the injector is from improper handling of the product.This injector has been in service for approximately 9 mos.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON MONARCH IV IOL DELIVERY SYSTEM, INJECTOR
Type of Device
FOLDERS AND INJECTORS , INRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18281682
MDR Text Key329901167
Report Number2523835-2023-00686
Device Sequence Number1
Product Code MSS
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K212039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIV
Device Catalogue Number8065977774
Device Lot Number14LMDA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received12/07/2023
Supplement Dates Manufacturer Received01/03/2024
02/17/2024
Supplement Dates FDA Received01/25/2024
03/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/22/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-