|
Model Number 48-20-00 |
Device Problem
Decreased Pump Speed (1500)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that during procedure speed of a cp5 console dropped to 1000 rpm, which is the minimum threshold with no alarm showing.Speed was then tried to be increased, however there was no response from the cp5 control panel.System was rebooted and the issue was still present.There was no patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.There was no patient involvement.H10: livanova deutschland manufactures the cp5 console.The incident occurred in houston, texas.Through follow-up communication livanova learned that pressure threshold was set to zero by the customer.Therefore, when the pressure went below zero it would start to ramp down the rpm and not alarm.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Manufacturer Narrative
|
Additional information has been requested and it has been revealed that the cp5 acts as per expectation and no alarm was requested based on additional information confirmed by the field service engineer, the event has been re-evaluated as not reportable event as the device acts as expected.
|
|
Search Alerts/Recalls
|
|
|