LIVANOVA USA, INC. LEAD MODEL 304 Back to Search Results
Model Number 304-20
Device Problems Fracture (1260); Low impedance (2285); Material Twisted/Bent (2981); Naturally Worn (2988)
Patient Problem Convulsion, Clonic (2222)
Event Date 11/07/2022
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that the patient has high impedance.Imaging also revealed a possible fracture.It was noted that the patient was having grand mal seizures and up until that time he had been seizure free since his implant with only occasional absence seizures.Device history records were reviewed for the lead and showed the device passed all specifications prior to distribution.The patient underwent a full replacement.The explanted devices have not been received to analysis to date.
 
Event Description
The suspect device was returned and received into product analysis for device testing.Testing has not been completed to date.No other relevant information has been received to date.
 
Event Description
Device evaluation was completed on the returned devices.High impedance was observed sooner than previously reported.
 
Manufacturer Narrative
Medical device problem :a072102.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 304
Type of Device
LEAD
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key18283597
MDR Text Key329950106
Report Number1644487-2023-01750
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number304-20
Device Lot Number204419
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received12/07/2023
Supplement Dates Manufacturer Received01/17/2024
03/04/2024
Supplement Dates FDA Received02/09/2024
03/13/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age6 YR
Patient SexMale
-
-