Device evaluated by mfr: the device was returned for evaluation.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually examined for damage or any irregularities.The catheter shaft showed multiple bends and kinks.Functional testing was attempted; however, the device would not prime.Damage was observed measuring from the proximal marker band to the tip at 4 cm, the entire section was stretched.During analysis at the severely kinked location at 1.5 cm from the tip to the proximal marker band, the hypotube was broken and completely separated/missing.The device could not be functionally tested due to the extreme damage on the device.No pressure reading was reported before the console would stop the priming process and issue an under-pressure alarm.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|