Related manufacturing reference: 3005334138-2023-00557.During an atrial fibrillation procedure, a pericardial effusion occurred requiring a pericardiocentesis to stabilize the patient.When the ablation catheter was inserted into the patient for the first time, the default sheath filter on ensite x version 3.0 did not recognize the catheter as "out of the sheath", despite all four of the electrodes being out of the sheath as confirmed by intracardiac echocardiogram, impedance and electrograms.The sheath filter needed to be manually reset in order to visualize the catheter.Shortly after the transseptal puncture, a blood pressure drop was noted, and a pericardial effusion was confirmed around the right ventricle via intracardiac echocardiogram.The procedure was stopped, a pericardiocentesis was performed and the patient recovered with no further consequences.The likely cause was an instance of high contact force in the left atrial appendage with the ablation catheter.Only 7 rf lesions were delivered prior to abortion of the ablation procedure, in which no abnormalities (high force or impedance/temperature rises) were noted.There are no reported performance issues with the ablation catheter.
|
The case study and collect logs were provided and reviewed.From the review of the files, it is observed that the catheter was momentarily outside of the sheath but went back in the sheath while attempting to collect the sheath baseline, leading to the sheath baseline not being collected.The sheath filter set baseline requires that all electrodes be out of the sheath.See ensite x ep system ifu, sheath filter.The system was operating as designed.
|