• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EMBLEM S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EMBLEM S-ICD; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT) Back to Search Results
Model Number A209
Device Problems Failure to Sense (1559); Under-Sensing (1661); Connection Problem (2900); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Event Description
It was reported that during the implant procedure there was minimal to no sensing in secondary or alternative vectors during auto setup.It was determined that the electrode was under-inserted in header with no contact with proximal electrode.Three different physicians were not successful at inserting the electrode further into the header even after multiple troubleshooting attempts.Subsequently the s-icd was attempted and not implanted due to concerns over an issue with the header.A new s-icd was implanted with the same electrode, and it was noted the electrode was able to be easily inserted into the header and no further connection issues were observed.
 
Manufacturer Narrative
No additional information is available.If additional information becomes available the report will be updated at that time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMBLEM S-ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key18287648
MDR Text Key330268092
Report Number2124215-2023-69677
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberA209
Device Catalogue NumberA209
Device Lot Number191376
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/13/2023
Initial Date FDA Received12/07/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/09/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-