The customer reported to olympus that during routine microbiological testing, the cystonephrofiberscope tested positive for 20 colony forming units (cfus)/100ml of contamination.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
The suspect device has been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The cleaning, disinfection, and sterilization (cds) was performed by the customer.The customer confirmed that there were no deviations or deficiencies concerning reprocessing of the scope.Additionally, the customer confirmed that there were no suspected patient infections due to the facility findings.The device was reprocessed 6 times prior to sampling, which was performed immediately after reprocessing.The device was quarantined following the positive microbiological test results and was not used for any procedures.The customer provided the cleaning, disinfection, and sterilization process stating that if manual cleaning was not performed within one hour after the procedure, pre-soaking was performed.During manual cleaning, a passing leak test was performed.The detergent used was franklab ddn9 and the device was rinsed before manual disinfection.The disinfectant used was franklab peralex 9 hecto plus and the channels were flushed and immersed with disinfectant, and rinsed with sterile water.The concentration and expiration date of the disinfectant were controlled.The device was stored in a simple cabinet.Olympus is the maintenance company.The hygiene microbiological investigation report indicated the channels of the scope were cultured and testing detected <1 colony forming units (cfus)/100ml of contamination.The results obtained comply with the target level for an endoscope subjected to high level disinfection and rinsed with sterile water.The device evaluation found the biopsy channel mount was damaged.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.The instruction manual includes the reprocessing method as is described in the following sections.This may prevent the phenomenon.- chapter 4 reprocessing workflow for endoscopes and accessories.- chapter 5 reprocessing the endoscope.Olympus will continue to monitor field performance for this device.
|