C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 9808560
Device Problems Fracture (1260); Material Separation (1562); Improper or Incorrect Procedure or Method (2017); Deformation Due to Compressive Stress (2889); Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Event Description
It was reported that approximately four years after port placement, the catheter was allegedly broken.Reportedly, the port was removed and replaced.There was no reported patient injury.
 
Manufacturer Narrative
H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
 
Event Description
It was reported that approximately four years after port placement, the catheter was allegedly broken.Reportedly, the port was removed and replaced.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one powerport isp mri implantable port attached to a groshong catheter in two segments was returned for evaluation.Visual, microscopic, functional and tactile evaluations were performed on the returned device.A complete circumferential break was noted on the distal end of the attached catheter and on the proximal end of the distal catheter segment.The edges of the complete circumferential breaks were noted to be uneven and the surfaces were noted to be round and smooth on one region and granular in the other region.Therefore the investigation is confirmed for the reported fracture and the identified material separation, deformation and material wear issues.The investigation is also confirmed for the identified improper procedure issue as the port was used after four years with no regular flushing.The definitive root cause for the identified improper procedure issue is use related.It is unknown whether patient and/or procedural issues contributed to the reported event.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.The instruction for use states that flushing volumes: when port not in use- 5 ml heparinized saline every 4 weeks.H10: g3, h6 (device) h11: h6 (method, result, conclusion) h11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT ISP M.R.I. IMPLANTABLE PORT, GROSHONG SINGLE-LUMEN, 8F
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18289582
MDR Text Key329994400
Report Number3006260740-2023-05591
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741027512
UDI-Public(01)00801741027512
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9808560
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received02/28/2024
Supplement Dates FDA Received03/08/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-