Model Number DM4500
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the implantable cardiac monitor (icm) had failed to be interrogated prior to procedure.No intervention was reported.There was no patient involvement.
 
Manufacturer Narrative
Customer complaint of failure to interrogate was confirmed.The device was received from the field with battery voltage having reached end of life.Device could not establish telemetry and device image could not be saved.Longevity assessment was performed, and device was found to have premature discharge of battery.Device was cut open and high drain current on the hybrid was noted.An anomalous integrated circuit (ic) was found to be the cause of the high current drain and failure to interrogate.A device history record (dhr) review was performed, and all required manufacturing processes and inspections steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOT DX¿ ICM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18290581
MDR Text Key330041211
Report Number2017865-2023-94261
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM4500
Device Lot NumberS000089276
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/19/2023
Initial Date FDA Received12/08/2023
Supplement Dates Manufacturer Received02/09/2024
Supplement Dates FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-