Catalog Number D142901
Device Problem Coagulation in Device or Device Ingredient (1096)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/13/2023
Event Type  malfunction  
Manufacturer Narrative
The device has been reported as discarded, therefore no product investigation can be performed, and the customer complaint cannot be confirmed.Manufacturing record evaluation (mre) cannot be conducted because no lot number was provided by the customer.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial flutter (afl) procedure with a optrell mapping catheter with trueref technology and a clot issue occurred.The catheter was in the body for 10 minutes.After the case, the catheter was removed and there was a clot lodged between the splines.The reporter wanted documentation of the issue.Patient asymptomatic at this time.Additional information was received.The clot was on the tip electrode.The system did not present any error messages nor did the physician/user see any product problem.There were no issues related to flow on the catheter.The patient was not anticoagulated.Catheter was used in the right atrium.The patient has not exhibited any neurological symptoms since the procedure was completed.The physician considered the clot was excessive based on their clinical experience.The physician did not consider the amount of clot observed caused a potential risk to this patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTRELL MAPPING CATHETER WITH TRUEREF TECHNOLOGY
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18292005
MDR Text Key330054555
Report Number2029046-2023-02884
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K230253
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD142901
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/13/2023
Initial Date FDA Received12/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM; SMARTABLATE GENERATOR KIT-US; SMARTABLATE PUMP KIT-US
-
-