• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON Back to Search Results
Model Number 97800
Device Problem Failure to Deliver Energy (1211)
Patient Problems Fecal Incontinence (4571); Urinary Incontinence (4572)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
B3: date is approximate.Month and year are confirmed valid.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for urinary/bowel disfunction and urge incontinence.It was reported that the patient stated that they were put on a diuretic by their managing physician for over a year and they 'upped it' last month and was possibly attributing the 'upping' to the problems they were encountering relating to their symptoms.The patient stated that starting a month ago they noticed a lack of bladder control.The patient stated they noticed a lack of bowel control yesterday.The caller stated they had been waiting to see their clinician but their 'community care' was causing them roadblocks with seeing their doctor.The caller went on to say that they believe their ins had turned itself off.The caller stated they checked the ins today and saw it was off but they didn't recall turning it off though they did note that they had been changing programs, upping their stimulation in an attempt to address their symptoms without any success.Caller stated they were tu rning stimulation up high without symptom relief.Agent advised that the ins should not turn itself off unless prompted by the external components and agent advised that the best course of action is to follow up with managing doc's office to see if insurance issue can be resolved so patient can subsequently be seen to have system/programming evaluated as well as a look at clinician data in reference to their concern around the ins possibly turning itself off.Patient did note that they has had multiple systems and that they have had success with the interstim in the past and it has given them a life.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM X
Type of Device
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key18292431
MDR Text Key330133435
Report Number3004209178-2023-25435
Device Sequence Number1
Product Code EZW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number97800
Device Catalogue Number97800
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received12/08/2023
Date Device Manufactured05/17/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age61 YR
Patient SexFemale
-
-