ALCON RESEARCH, LLC - HUNTINGTON MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL) Back to Search Results
Catalog Number 8065977763
Device Problem Device Damaged by Another Device (2915)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/09/2023
Event Type  malfunction  
Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event Description
A non healthcare professional reported that a crack was found in the optic (near the root of trailing haptic) after intraocular lens (iol) was implanted.The surgery was completed without product replacement, and there's no effect on the postoperative va.The iol was still remains in the patient's eye.Additional information has been requested.
 
Manufacturer Narrative
Two used company iii (d) cartridges were returned labeled for two files.There is no way to determine which cartridge goes to which file.The cartridges were numbered 1-2 for evaluation purposes.Both returned used company iii (d) cartridges were microscopically examined.The first cartridge did not have any damage.The second used company iii (d) cartridge did not exhibit adequate viscoelastic.This cartridge had heavy stress and an aneurysm on the left side of the tip.Both cartridges had evidence of placement into a handpiece.Top coat dye stain testing was conducted with acceptable results.Product history records were reviewed and documentation indicated the product met release criteria.Qualified associated products were indicated.The reported lens damage could not be verified.The lens remains implanted and a photo was not provided.The root cause for the reported lens damage may be related to a failure to follow the instructions ifu.Both cartridges were top coat dye stain tested with acceptable results.One cartridge exhibited inadequate viscoelastic.This cartridge tip had heavy stress and an aneurysm.This type of damage may occur if the lens is not positioned correctly for advancement; if there is a lack of viscoelastic between the lens and the cartridge lumen; if the lens is advanced too rapidly or if the handpiece plunger is not positioned correctly at the trailing optic edge.If the handpiece plunger is not positioned at the trailing optic edge, it can allow the lens to fold around the plunger tip making it too large to correctly advance through the narrow tip of the cartridge, which could cause damage to the tip or the lens.The company cartridge ifu instructs to completely fill the cartridge with ophthalmic viscoelastic device (ovd) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.The company handpiece ifu instructs: important: the plunger should make initial contact with the cartridge at the ramp.In the event the plunger does not contact the cartridge at the ramp, do not use the handpiece and contact company.Verify that the plunger tip is properly engaging the lens optic and that the lens moves forward at the same rate as the plunger while slowly advancing the plunger forward to avoid damaging the lens.When the threads on the knob make contact with the barrel, turn the knob clockwise approximately ½ turn to engage the threads and then stop.The iol will now be in the dwell position.Inspect to ensure the plunger is behind the optic.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D
Type of Device
FOLDERS AND INJECTORS, INTRAOCULAR LENS (IOL)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18299780
MDR Text Key330113388
Report Number1119421-2023-02114
Device Sequence Number1
Product Code MSS
UDI-Device Identifier00380659777639
UDI-Public00380659777639
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065977763
Device Lot Number15664884
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/16/2023
Initial Date FDA Received12/11/2023
Supplement Dates Manufacturer Received03/11/2024
Supplement Dates FDA Received04/04/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2023
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CLAREON MONARCH IV INJECTOR; CLAREON PANOPTIX TORIC TRIFOCAL HYDROPHOBIC IOL; PROVISC
-
-