• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ROTAPRO; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number 39467-150
Device Problems Mechanical Problem (1384); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/21/2023
Event Type  malfunction  
Event Description
It was reported that the burr detached.The target lesion was located in a mildly tortuous and moderately calcified right coronary artery (rca).A 1.50mm rotapro was prepared and attempted to be inserted, but the wire moved distally because the brake was not defeated while the burr was spinning upon advancing into the guide (due to improper technique).Subsequently, the rotapro was removed and a non- boston scientific atherectomy device was attempted but it also had difficulty.Then, the same burr was selected to be attempted again.However, while platforming to 170k rpm, the burr separated from the drive shaft while on the rotawire outside the patient.The burr was then taken off the wire and the case were rescheduled for a later day.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAPRO
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18300345
MDR Text Key330160457
Report Number2124215-2023-69415
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729893356
UDI-Public8714729893356
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/11/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number39467-150
Device Catalogue Number39467-150
Device Lot Number0032036408
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/22/2023
Initial Date FDA Received12/11/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ROTAWIRE GUIDEWIRE - BOSTON SCIENTIFIC
-
-