DEPUY SYNTHES PRODUCTS LLC MINIMAL ACCESS ATTACHMENT; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
|
Back to Search Results |
|
Catalog Number MA-D20 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/05/2023 |
Event Type
malfunction
|
Event Description
|
It was reported from japan that the screw dropped off from the motor device and attachment device.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure or if a spare device was available.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: d10: concomitant med products and therapy dates: motor device, 12/5/2023.E1: the reporter¿s phone number was not provided.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.Udi: (b)(4).
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed, and it was determined that the reported condition of the device falling apart ¿ unattached was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation, it was determined that the device had cosmetic damage, worn bearing, vibration and made an unexpected noise.It was further determined that the device failed pretest for visual assessment.The assignable root cause was determined to be due to component failure from normal wear.
|
|
Search Alerts/Recalls
|
|
|