Catalog Number LXMC15 |
Device Problems
Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
Pyrosis/Heartburn (1883)
|
Event Date 01/01/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Date sent 12/11/2023.B3: only event year known: 2023.D6a: exact date is unk.Assumed first month of the year.No lot number was provided therefore a device history could not be done.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the exact date of implant? when did the acid reflux symptoms begin? what was the date of the imaging which showed the discontinuous linx? if available, please share a copy of this imaging.Please send to: productcompliant1@its.Jnj.Com what is the product code? what is the device lot number? was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? when and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? when and if the linx device is removed, may we ask that the device be returned for analysis? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that during an x-ray, they found the linx device to be broken/discontinuous.Patient has had no mri with the device.Patient has mild reflux disease but not enough to where they needed to come in to have it looked at.Linx was implanted (b)(6) 2020.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 1/16/2024.Photo analysis: an x-ray image of the device in vivo was reviewed by a medical safety officer.As per medical safety officer: "the image reviewed demonstrates a discontinuous linx device located below the diaphragm.".The mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.Additional information received: what was the exact date of implant? january 2020; when did the acid reflux symptoms begin? july 2023; what was the date of the imaging which showed the discontinuous linx? if available, please share a copy of this imaging.Please send to: (b)(4), 11/28/23.What is the product code? not sure ¿ appears to be lxmc15.What is the device lot number? unknown.Was the device initially effective in controlling reflux? yes.Were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? epigastric discomfort.Did the patient undergo an mri since device implant? if so, when was the mri and what strength? no.Did the patient have any other surgeries in the area? none reported.Was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.None reported.What is the management plan? is device removal scheduled? is a replacement linx or fundoplication planned? egd & bravo study planned for (b)(6) 2024 / removal not scheduled / not sure.When and if the explanation takes place can we ask that the procedure gets video recorded and the video shared? i will inquire.When and if the linx device is removed, may we ask that the device be returned for analysis? yes d1, d4.
|
|
Search Alerts/Recalls
|
|
|